Infant Incubator DIXION BabyGuard I-1120 Transformer

Infant Incubator Mode (Close Mode)

For the provision of a similar maternal uterus environment  for the neonate to provide stable and controlled  comfortable thermal environment and servo-controlled  humidity and oxygen level. It is ideal for use in Infant temperature recovery, resuscitation and other occasions.

Infant Radiant Warmer Mode (Open Mode)

Uniform heat distribution, maintaining an optimal  thermal environment for caregiver to provide immediate access to the newborn, it is ideal for use in labor and delivery rooms, NICU, and Operating Room.

Monitor

Physical parameters were monitored such as Neonatal ecg, respiration, noninvasive blood pressure, oxygen saturation, pulse rate, Exhaled carbon dioxide.

Incubator Mode(Close Mode)

 

Mode of control Air mode and Baby mode controlled by

micro-computer

 

Air temperature control range 20°C-37°C(override mode 37°C-39°C)
Baby temperature control range 34°C-37°C(override mode 37°C-38°C)
Accuracy of skin temperature sensor within ± 0.2°C
The temperature fluctuation between incubator

temperature and average incubator temperature (under stable

temperature condition)

≤0.5°C

 

Temperature uniformity ≤0.8°C(mattress is on the horizontal position)

≤1.0°C(mattress is on the maximum tilt position)

 

Inclination of the bassinet ± 12° (stepless adjustable)
Noise inside hood (stable working condition) ≤50dB(A)
Relative Humidity Level:

Humidity display range

Humidity control range

Accuracy of humidity display

Accuracy of humidity control

 

0%RH~100%RH

0%RH~90%RH

± 5%RH

± 5%R

 

Oxygen Level:

Indication range of oxygen concentration

Control range of oxygen concentration

Indication accuracy of oxygen concentration

 

Control accuracy of oxygen concentration

 

0%~100%

20%~60%

± 2% (The set value is below 25%)

± 3% (The set value is above 25%)

±4% 02

Infant Radiant Warmer Mode(Open Mode)

 

Control Mode Pre-warm mode, manual mode and baby mode

are controlled by micro-computer

Baby mode temperature control range 34°C-38°C
Control accuracy ( The fluctuation range between Skin temperature and Set temperature) ≤0.5°C

 

Temperature uniformity of mattress ≤2°C

 

Performance Index Under Thepulse Oximetry

 

SPO2 Display range 1- 100%
SPO2 Display resolution 1%
SPO2 Alarm setting range Upper limit:50 ~100%

Lower limit: 45-95%

 

PR Display range 25-240 bpm
PR Display resolution 1bpm
PR Alarm Setting range Upper limit: 80 – 240 bpm

Lower limit: 35-180 bpm

PI Display range 0.02-20.0%
PI Display Resolution 0.01%
Sensitivity setting Normal,APOD, Max,

Factory default setting: Normal

FastSAT off, on, Factory default setting: off
Mean time 2S -4S, 4S-6S, 8S,10S, 12S, 14S, 16S

Factory default setting: 8S

FailureAlarm

 

Power Failure Alarm, Fan Motor Alarm, Sensor, FailureAlarm, Deviation Alarm, CanopyFailure Alarm, OverTemperature Alarm, Water tank placed false alarm, WaterShortage Alarm, SettingAlarnr Checking Alarm, System _

Alarm, etc.

Transportation

 

Storage Temperature range: -20°C~+55°C

Relative Humidity range: ≤93%

Atmospheric pressure: 500hPa – 1060hPa

Environment Requirement

 

Operating range: 20°C-30°C(Set temperature

should be 3°C higher than the environment temperature)

Relative Humidity range: 30%~75%

Ambient air velocity rate: ≤0.3m/s

Product Packaging

 

Package: The whole unit is packed into one carton

Size: 1140mm x 660mm x 2315mm

Gross Weight: 160Kg

Power supply

Power Requirement

Auxiliary Mains Power Output

Current

AC220V-230V/50Hz

1300VA

AC220-230V/50Hz, Max

1A

Standard Configuration

Main body (including baby compartment, Base, Controller,

I.V.pole and tray), Sensor box, Skin temperature sensor and

Vertical height adjustment stand (VHA stand), Warmer module.

Optional Configuration

Weighing system, Monitor shelf and Disposable

skin temperature sensor, Observation lamp, Sp02, etc.