Infant Incubator Mode (Close Mode)
For the provision of a similar maternal uterus environment for the neonate to provide stable and controlled comfortable thermal environment and servo-controlled humidity and oxygen level. It is ideal for use in Infant temperature recovery, resuscitation and other occasions.
Infant Radiant Warmer Mode (Open Mode)
Uniform heat distribution, maintaining an optimal thermal environment for caregiver to provide immediate access to the newborn, it is ideal for use in labor and delivery rooms, NICU, and Operating Room.
Monitor
Physical parameters were monitored such as Neonatal ecg, respiration, noninvasive blood pressure, oxygen saturation, pulse rate, Exhaled carbon dioxide.
Incubator Mode(Close Mode)
|
|
Mode of control | Air mode and Baby mode controlled by
micro-computer
|
Air temperature control range | 20°C-37°C(override mode 37°C-39°C) |
Baby temperature control range | 34°C-37°C(override mode 37°C-38°C) |
Accuracy of skin temperature sensor | within ± 0.2°C |
The temperature fluctuation between incubator
temperature and average incubator temperature (under stable temperature condition) |
≤0.5°C
|
Temperature uniformity | ≤0.8°C(mattress is on the horizontal position)
≤1.0°C(mattress is on the maximum tilt position)
|
Inclination of the bassinet | ± 12° (stepless adjustable) |
Noise inside hood (stable working condition) | ≤50dB(A) |
Relative Humidity Level:
Humidity display range Humidity control range Accuracy of humidity display Accuracy of humidity control |
0%RH~100%RH 0%RH~90%RH ± 5%RH ± 5%R
|
Oxygen Level:
Indication range of oxygen concentration Control range of oxygen concentration Indication accuracy of oxygen concentration
Control accuracy of oxygen concentration |
0%~100% 20%~60% ± 2% (The set value is below 25%) ± 3% (The set value is above 25%) ±4% 02 |
Infant Radiant Warmer Mode(Open Mode)
|
|
Control Mode | Pre-warm mode, manual mode and baby mode
are controlled by micro-computer |
Baby mode temperature control range | 34°C-38°C |
Control accuracy ( The fluctuation range between Skin temperature and Set temperature) | ≤0.5°C
|
Temperature uniformity of mattress | ≤2°C
|
Performance Index Under Thepulse Oximetry
|
|
SPO2 Display range | 1- 100% |
SPO2 Display resolution | 1% |
SPO2 Alarm setting range | Upper limit:50 ~100%
Lower limit: 45-95%
|
PR Display range | 25-240 bpm |
PR Display resolution | 1bpm |
PR Alarm Setting range | Upper limit: 80 – 240 bpm
Lower limit: 35-180 bpm |
PI Display range | 0.02-20.0% |
PI Display Resolution | 0.01% |
Sensitivity setting | Normal,APOD, Max,
Factory default setting: Normal |
FastSAT | off, on, Factory default setting: off |
Mean time | 2S -4S, 4S-6S, 8S,10S, 12S, 14S, 16S
Factory default setting: 8S |
FailureAlarm
|
Power Failure Alarm, Fan Motor Alarm, Sensor, FailureAlarm, Deviation Alarm, CanopyFailure Alarm, OverTemperature Alarm, Water tank placed false alarm, WaterShortage Alarm, SettingAlarnr Checking Alarm, System _
Alarm, etc. |
Transportation
|
Storage Temperature range: -20°C~+55°C
Relative Humidity range: ≤93% Atmospheric pressure: 500hPa – 1060hPa |
Environment Requirement
|
Operating range: 20°C-30°C(Set temperature
should be 3°C higher than the environment temperature) Relative Humidity range: 30%~75% Ambient air velocity rate: ≤0.3m/s |
Product Packaging
|
Package: The whole unit is packed into one carton
Size: 1140mm x 660mm x 2315mm Gross Weight: 160Kg |
Power supply
Power Requirement Auxiliary Mains Power Output Current |
AC220V-230V/50Hz
1300VA AC220-230V/50Hz, Max 1A |
Standard Configuration
Main body (including baby compartment, Base, Controller,
I.V.pole and tray), Sensor box, Skin temperature sensor and
Vertical height adjustment stand (VHA stand), Warmer module.
Optional Configuration
Weighing system, Monitor shelf and Disposable
skin temperature sensor, Observation lamp, Sp02, etc.