Patient Monitor DIXION Storm 5120

  • Apply to hospital, clinic and home
  • Suitable for adult, pediatric and neonate
  • Portable, easy to carry
  • 4.3”color TFT LCD Touch screen with auto rotation function
  • Minimal warm-up time for EtCO2 and RR measurements
  • CO2 water filter ensuring extended monitoring in a humid environment
  • Selectable layout modes including font and waveforms
  • Dual work modes: continuous monitoring & spot check
  • Adjustable audio and visual alarms
  • Large data storage capacity up to 500 groups per ID
  • Large capacity Lithium battery can work for more than 18 hours with full charge
  • Automatic power-off function for power saving
Mainframe size 142mm(W)×78mm(H)×36mm(D)
Mainframe weight <500g (full configuration,including  batteries)

 

Power adapter Input AC 100 – 240V
Type Built-in lithium battery

Voltage: 3.7V

capacity 4800mAH

Charging time: 3 hours to 90%, 4 hours to 100%

Run time: >18h

 

Display Size: 4.3 inch

Resolution: 480×272

Data storage Displaying way: Trend tabular

SpO2 trend interval: 10s, 30s, 1min, 2min, 5min, 10min

Capacity: 1-96 patient ID, 500 groups/patient(only data)

Spot-check: ID from 1 to 99, 300 groups can be stored for each ID

Continuous: ID is 0, 60000 groups can be stored.

CO(Sidestream) Measurement Way: Infrared spectrum

Measurement Range: 0—19.7%( 0-150mmHg or 0-20.0kPa

CO2 Accuracy: 0%-5.3%(0mmHg-40mmHg),±0.3%(±2mmHg);

5.4%-9.2%(41mmHg-70mmHg), ±5% of reading;

9.3%-13.2%(71mmHg-100mmHg);±8% of reading;

13.3%-19.7%(101mmHg-150mmHg), ±10% of reading.

CO2 Resolution: 0.1mmHg

Sample Flow Rate: 60~80ml/min

Unit: %, mmHg, kPa

Measurement Range of awRR: 3~150 rpm

Measuring accuracy of RR: ±1% or ± 1 rpm,

whichever is greater

Response Time:<3s

SpO2 Measurement range: 0~100%

Resolution: 1%

Accuracy: 70~100%

 

Standard configuration: ETCO2, RR, Multi-parameter Display Mode, CO2 Waveform Display Mode,

SpO2 Display Mode

Option:SPO2, PR,  Trolley, Silicone case

Basic service