- Modern complete digital design
- Foldable antiglare LCD display
- 12 leads simultaneous acquisition
- High resolution thermal printer
- Built-in rechrgable Li-ion battery
- Well designed touch pannel
- RS-232 port
- Automatic adjustment of baseline for optimal recording
- Complete digital filter, avoiding baseline drift, AC, EMG interference
- Baseline automatic adjusment, printing position
- Auto/manual operation
- Lead-off detection and alarm
- Selectable printing formats, standard single channel or single channel & rhythm lead
Specification:
| Leads | 12 standard leads |
| Input Circuit | Floating input circuit, protection from defibrillation |
| Time constant | ≥3.2s |
| Frequency Response | 0.05Hs – 150 Hz |
| Sensitivity | 2.5; 5; 10; 20 (mm/mV), AGC |
| Input Impedance | ≥50MΩ |
| Input Circuit Current | ≤50 nA |
| Noise Level | ≤12.5μVp-p |
| Anti Baseline Drift | Automatic |
| Filter | EMG-Filter: 35 Hz (-3dB)
AC/DFT-Filter 50 Hz/60Hz (-20 dB) |
| CMRR | >110dB |
| Recording Mode | Auto/Manual |
| Safety Standarts | IEC 60601-1:2005 EN 60601-1:2006 EN 60601-1-2:2007 IEC 60601-1-2:2007 IEC/EN60601-2-25 |
| Power Supply | AC: 100 V-115 V / 220 V-240 V, 50/60 Hz/ Li-ion battery: voltage 14.8 V More than 6.5 h |
| Recorder | Thermal printer
Dot structure: 384 dots/line Dot pitch: 0.125 mm (8 dots/mm) Dot size: 0.125mmx0.12mm |
| Record paper | 50mm x 30 m (rolled thermal paper) |
| Paper speed | 25mm/s, 50mm/s |
| Dimension | 288x 210x 70 mm |
| Weight | 2,3kg (including battery) |
| Display | 192×64 pixels LCD-interface |
| External Input/Output | Input: ≥ 100kΩ; Sensitivity 10mm/V ± 5%
Output: ≤ 100kΩ; Sensitivity 1V/mV±5% |
| Communication | RS-232 |