Safe, stable and efficient anesthesia management.
- The characteristic breathing circuit is made of alloy, resistant to corrosion and can withstand repeated high temperature and high pressure sterilization.
- Adjustable angle, easy to install, many user-friendly designs make maintenance easier.
- The integrated heating system with a better thermal conductivity of alloy help prevent condensation and make patients feel more comfortable.
- APL with fast release pressure, the upper pressure limit is accurately adjustable, avoiding repeated operations and improving anesthesia efficiency.
- The Breathing Ciucuit has CO2 bypass function.
- Practice 3800A has a minimum tidal volume of 10ml in volume control mode, in addition to possessing the PCV-VG and BIVENT ventilation mode, helping to achieve the precise low tidal volume required during lung protective ventilation.
Individualized PEEP titration tool
- Stress index (SI) monitoring helps with Individualized PEEP titration. Through the guidance of the Static PV loop tool, the appropriate setting of PEEP value and tidal volume are realized
Digital Flowmeter with ECO-Optimizer
- Digital Flowmeter makes fresh gas flow setting easier and more precise.
- The ECO-Optimizer indicates the recommended fresh gas flow setting according to the setting value and the minimum O2 needed of the patient. It enables a safe Low Flow, and reduces the waste of anesthetic agents and medical gases
Minimized impact recruitment maneuver
- Two types recruitment maneuvers: stepwise PEEP or sustained inflation.
- Automate repetitive tasks used during lung ventilation procedures.
- Enhanced monitoring and clinical tools
- In addition to traditional monitoring parameters, special monitoring parameters, such as Driving Pressure(DP), are provided to guide clinicians in adjusting ventilation parameters.
- Spirometry loops can be stored for future reference, a [lowing clinicians the ability to better understand changes in the patient’ s response to therapy.
- Provide multiple of cardiopulmonary bypass modes (CBP) to assist in the implementation of cardiopulmonary bypass surgery.
- Continuous trend information together with time discrete events are stored and shown in the table or chart.
- Provides medical gas consumption calculations: including O2, NzO and Agent. And provide calculations of CO2 production.
BASE UNIT | |
Dimensions (H x W x D)
Trolley version (with breathing circuit) |
1420mm×770mm×760mm |
Weight :
Trolley (without vaporizer and backup cylinder) Top shelf load |
135kg 25kg |
Power input | 100~240 VAC, 50/60 Hz |
Batteries and Operation time with fully charged | DC24V, 4.0AH, Min.120min |
Operation temperature | 10~+40ºC |
Operation humidity | ≤95%(non-condensing) |
ANESTHESIA GAS SUPPLY MODULE | |
Gas supply | O2, N2O, AIR 280kPa – 600kPa |
Cylinder yokes Optional | O2, N2O, AIR |
Fresh gas flow indicator | Electronically display and virtual fresh gas flow tubes |
Range of fresh gas flow indicators | O2, N2O (0~10.0 l/min), AIR (0~12.0 l/min)
O2flush 25~75 l/min |
ACGO | Optional |
AGSS | Optional |
VAPORIZER | |
Agent | Halothane, Enflurane, Isoflurane , Sevoflurane |
Installation mode | Selectatec® with interlock, optional standby vaporizer parking holder |
Filling type | Pour Fill, Key Fill, Quik-Fil |
APL Range | Spontaneous breathing (SP) -70cm H2O |
CO2bypass | Optional |
VENTILATOR OPERATING SPECIFICATIONS | |
Ventilator | Electronically controlled, pneumatically driven |
Ventilation modes | Manual/spontaneous
IPPV, PCV, PCV-VG, SIMV-VC, SIMV-PC, SIMV-VG, CPAP,BIVENT APRV |
CONTROL INPUT RANGES | |
Breathing frequency (rate) | 2~100bpm |
PEEP | OFF, 3~50cm H2O |
Inspiration/expiration ratio (Ti:Te) | 4:1~1:8 |
Tidal volume | 20~1500ml(10~1500ml by Volume Control and SIMV modes)(5 to 1500 ml by PressureControl Vent Mode) |
Inspiration pause | OFF, 5%~60% |
Inspiratory time | 0.2~5.0 s |
Inspiratory pressure | 5~70cm H2O |
Pressure support level | 3~60cm H2O |
Pmax | 10~100 cm H2O |
Trigger level | 0.5~15 l/min |
Monitoring |
Continuous monitoring of inspiratory O2concentration, breathing frequency, tidal volume, minute volume, peak airway pressure, PEEP, mean or plateau pressure, CO2 concentration (optional), anesthetic gas concentration (optional) paramagnetic oxygen sensors (optional) |
Compliance | 0~100 mL/cm H2O |
Control screen | 15 inch; TFT color touch screen |
Graph Display | Waveform of P-t, F-t, V-t, CO2-t (option), Paw-V Loop, V-Flow Loop,P-F Loop |
Alarm |
MV high/low limit, FiO2high/low limit, Paw high/low limit, Power failure Breath Rate high limit
Subatmospheric pressure Sustained airway pressure Apnea alarm, Alarm Silence(120s)
|